| Summary | There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure. |
|---|---|
| Publication Date | Nov. 15, 2024, 6:15 a.m. |
| Registration Date | Nov. 15, 2024, noon |
| Last Update | Nov. 15, 2024, 10:58 p.m. |