NVD Vulnerability Detail
Search Exploit, PoC
CVE-2019-25719
Summary

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Publication Date June 2, 2026, 11:16 p.m.
Registration Date June 3, 2026, 4:18 a.m.
Last Update June 2, 2026, 11:40 p.m.
CVSS3.1 : HIGH
スコア 8.6
Vector CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
攻撃元区分(AV) ネットワーク
攻撃条件の複雑さ(AC)
攻撃に必要な特権レベル(PR) 不要
利用者の関与(UI) 不要
影響の想定範囲(S) 変更なし
機密性への影響(C)
完全性への影響(I)
可用性への影響(A)